Ezetimibe and Simvastatin Tablets
Failed Impurities/Degradation Specifications: Out of Specification (OOS) for related substances test for Anhydro Simvastatin at the 06-month time point during long-term stability study.
Class II — Moderate risk. May cause temporary or reversible harm; serious harm unlikely.
Failed Excipient Specifications and Presence of Foreign Tablets/Capsules; product manufactured using an excipient found to be OOS for conductivity and some Ezetimibe and Simvastatin Tablets, 10 mg/10 mg were found in the bottle
Cause type: Particulate / foreign matter
Visible foreign particles - glass, metal, fibers or other material - found in the product. Particles can enter from equipment wear, containers or the process, and are dangerous in injectables. All particulate / foreign matter recalls.
Ezetimibe and Simvastatin Tablets, 10 mg/40mg, a) 90 Tablets, b) 500 tablets, Rx only, Manufactured by Dr. Reddy's Laboratories LA LLC Shreveport, LA 71106 USA, Packaged by GSMS, Incorporated Camarillo, CA 93012 USA; NDC: a) 51407-192-90, b) 51407-192-05
Compare with the code on your package — it is the most precise check available.
Drug: Simvastatin Category: Cholesterol Medication
From the FDA NDC directory, for the product code this recall names. Compare with the label on your package.
| Product | Strength | Form & route | Labeler | Type |
|---|---|---|---|---|
| Ezetimibe and Simvastatin NDC 51407-0192 |
EZETIMIBE 10 mg/1; SIMVASTATIN 40 mg/1 | Tablet Oral |
Golden State Medical Supply, Inc. | Rx |
Check, then ask. Class II means harm is possible but usually temporary or reversible. Find the lot number on your package; if it matches, ask your pharmacist about a replacement. Do not simply stop a prescribed medication on your own - the gap can be riskier than the recall. This recall is now terminated, which means the firm has finished removing the affected product; anything on pharmacy shelves today should be unaffected.
Failed Impurities/Degradation Specifications: Out of Specification (OOS) for related substances test for Anhydro Simvastatin at the 06-month time point during long-term stability study.
CGMP deviations.
CGMP deviations.
CGMP deviations.
cGMP Deviations for the manufacturing Firm (Accord Healthcare) after their inspection.
Failed Dissolution Specifications. Notification from the manufacturer that lot is being recalled due to failed dissolution
CGMP deviations: tablets with black specks.
Failed Dissolution Specifications: During 12-month long-term stability testing, subject lot was out of specification (low) for stage 3 dissolution at the 24-hour timepoint.
Cross contamination with other products: API contaminated with trace amounts of Acemetacin API.